Removal Procedure (cont'd)
2.) Disconnect one connector (11) from the main circuit board in the keyboard. Refer to Figure 8-43.
3.) Unscrew four (4) screws (12-15) from the LCD unit. Refer to Figure 8-44.
(12)
(13)
4.) Remove the LCD Assy from the Keyboard.
5.) Perform the following functional tests. If all are successful, include the debrief script provided below.
Table 8-18
Service Manual
Section 4-3-1
Section 4-3-2
Section 4-7-8
8-4-11-5
Mounting Procedure
Install the new parts in the reverse order of removal.
8-36
2
Figure 8-43 Disconnect one connector
LCD Unit
Figure 8-44 Unscrew 4 screws from the LCD unit
Functional Tests
Section
Functional Test / Diagnostic Test
OP Panel LCD Assy Validation
Section 8-4 - Keyboard Block
Power On/Boot Up
Power Off / Shutdown
LOGIQ™5 S
ERVICE
(11)
(14)
(15)
Debrief Script
"Service Manual, Direction
2300000, Rev 1+, Section 8-4-11. Equipment
passed all required tests and is ready for use. "