Air Filter Set

Purpose: This is a description on how to remove and replace the Air Filter Set.
8-6-7-1
Tools
Not Required.
8-6-7-2
Needed Manpower
1 person, 5 minutes + travel
8-6-7-3
Preparations
Shut Down the System and switch off the Main Breaker at the rear as described in section 4-3-2 on
page 3.
8-6-7-4
Removal Procedure
1.) Remove Air Filter as illustrated below.Refer to Figure 8-87.
2.) Perform the following functional tests. If all are successful, include the debrief script provided below.
Table 8-42
Service Manual
Section 4-3-1
Section 4-3-2
Section 4-8-2-7
Section 10-5-5
8-6-7-5
Mounting procedure
Install the new parts in the reverse order of removal.
8-88
2
Figure 8-87 Removing the Air Filters
Functional Tests
Section
Functional Test / Diagnostic Test
Section 8-6 - Body Block
Power On/Boot Up
Power Off / Shutdown
Air filter set
Physical Inspection
LOGIQ™5 S
ERVICE
Debrief Script
"Service Manual, Direction
2300000, Rev 1+, Section 8-6-7. Equipment
passed all required tests and is ready for use. "