GE M
S
EDICAL
YSTEMS
D
2300000, R
IRECTION
EVISION
6-3-1-4
Removal Procedure (cont'd)
4.) Unscrew four 4 screws and disconnect twenty (20) connectors from the front side of the LV unit
using long-nose gripper.
Screw
5.) Remove the LV unit.
1
Screw
Figure 6-2 Disconnect 20 connectors
Chapter 6 Service Adjustment
LOGIQ™ 5 S
M
ERVICE
Screw
20 Connectors
Screw
ANUAL
6 - 3