GE LOGIQ 5 Service Manual page 293

Hide thumbs Also See for LOGIQ 5:
Table of Contents

Advertisement

GE M
S
EDICAL
YSTEMS
D
2300000, R
IRECTION
EVISION
8-7-4-4
Removal Procedure (cont'd)
6.) Disconnect one connector (1) if removing the FEC Assy. Otherwise skip this step. Refer to Figure
8-95.
7.) Move the upper and lower stopper in the direction indicated by the arrow (1). Refer to Figure 8-96.
8.) Pull out the board in the direction indicated by the arrow (2). do not bend it.
2
SIG Board
Figure 8-95 Disconnecting Cable
(1)
(2)
Figure 8-96 Pulling Out the Board.
Chapter 8 Replacement Procedures
(1)
LOGIQ™5 S
M
ERVICE
ANUAL
(1)
8-97

Advertisement

Table of Contents
loading

Table of Contents