Hipot Start parameter
The Hipot Start parameter, found in the System Menu screen, controls whether the
Hipot test will run sequentially after the Ground Bond test or simultaneously with the
Ground Bond test. Please refer to section 4.1 Setup System for instructions on how
to set this parameter.
The Hipot Start parameter is only used when connecting the HYAMP III with an
Associated Research Hypot III. The parameter determines if the Hypot will run after
the HYAMP III has completed its test, GW. If the Hypot will run at the same time as
the HYAMP III, G + W.
The Hipot Start parameter directly controls the Start output relay. Setting the Hipot
start parameter to GW will momentarily close the start output relay at the end of the
Ground Bond test. Setting the Hipot start parameter to G + W will momentarily close
the start output relay at the start of the Ground Bond test.
42