4.2.3. Default System Parameters
Setup System
PLC Remote
Single Step
Alarm
Contrast
Results
Lock
Mem Lock
Hipot Start
Date Format
Cal Alert
Alert Date
4.2.4. Memory, Step, and Default System Parameter Restore
Press the two bottom soft keys and power the instrument at the same time, then press
the soft key next to the word "Yes". All memories and steps will be loaded with the
ACW default parameters and the system parameters will be set to the factory defaults.
The following System parameters will not be affected by the system initialization:
Setup System
Cal Alert
Cal Date
Cal Due
Alert
Date m/d/y
Time
4.3. Test Setup
1. From the Perform Tests screen, press the "Menu" soft key. The Main Menu
screen will now be displayed.
2. From the Main Menu screen, press the "Menu" soft key. Results, Test and
System selections will now be displayed.
3. From this screen, press the "Test" soft key. The Test Parameter Review screen
will now be displayed.
4. From this screen all relevant test parameters may be edited and saved. Refer to
section 4.3.4. Default Test Parameters for a complete list of test parameters.
28
Restoring the instrument's System parameters will overwrite
all memories and steps with ACW default parameters!
OFF
OFF
5
5
Last
OFF
OFF
GW
mdy (month, day, year)
ON
Cal Date + 11 months
No change
No change
No change
No change
No change
No change