6. CONNECTION OF REMOTE I/O

Two 9 pin D-type connectors mounted on the rear panel provide REMOTE-INPUT-
OUTPUT control and information. These connectors mate with a standard 9 pin D-
type subminiature connector provided by the user. The output mates to a male (plug)
connector while the input mates to a female (receptacle) connector. For best
performance, a shielded cable should be used. To avoid ground loops the shield
should not be grounded at both ends of the cable. Suggested AMP part numbers for
interconnecting to the Remote I/O are shown below:
205204-4
PLUG SHELL WITH GROUND INDENTS
205203-3
RECEPTACLE SHELL
745254-7
CRIMP SNAP-IN PIN CONTACT (for plug)
745253-7
CRIMP SNAP-IN SOCKET CONTACT (for receptacle)
745171-1
SHIELDED CABLE CLAMP (for either plug or receptacle)
747784-3
JACKSCREW SET (2)

6.1. Remote Signal Outputs

The rear panel connector provides three output signals to remotely monitor PASS,
FAIL, and PROCESSING conditions. The monitoring signals are provided by three
normally open internal relays that toggle ON and OFF to indicate the condition of the
tester. The RESET OUT pulse signal and the START OUT pulse signal are also
provided by a normally open internal relay. The RESET OUT gives a signal whenever
the RESET function is activated. This can be used to abort a Hipot test while the units
are interfaced as a test system. The START OUT gives a momentary output pulse
36