IKONIX Associated Research HYAMP III Operation And Service Manual page 39

Ground bond tester
Hide thumbs Also See for Associated Research HYAMP III:
Table of Contents

Advertisement

Abort: This message appears on the display, if the test in process is aborted with the
RESET button or remote Reset control.
Max-Fail: This message appears on the display if the measured reading exceeds the
Max Lmt current setting of the Ground Bond test, or Max Lmt resistance setting of the
Insulation Resistance test.
Min-Fail: This message appears on the display if the measured reading drops below
the Min Lmt current setting of the Ground Bond test, or Min Lmt resistance setting of
the Insulation Resistance test.
Short: This message appears on the display, if the DUT current is well beyond the
metering range of the test.
Interlock Open: This message appears on the display if the Remote Interlock feature
is activated before or during a test. See section 6.2. Remote Signal Inputs and
Memory Access for more information.
Out-Error: This message appears on the display if the instrument has an internal
problem and the TEST button is pressed. The RESET button is not active while this
screen is displayed. To clear this screen and return to test mode press the "Exit" soft
key. When the "Exit" soft key is pressed the instrument will continue with its normal
failure indication process. The failure light and Alarm can then be cleared by pressing
the RESET button.
NOTE: If the instrument continues to power up in this condition it should be
sent in for service or repair.