It is possible to quickly bypass the security of the Cal-Alert by powering up the
instrument while pressing the soft key that is second from the top of the instrument.
This will temporarily start the instrument in an unsecured state. The Cal-Alert Warning
screen will be displayed and Cal-Alert may be disabled by selecting "No" and then
pressing the "OK" soft key. This type of security override only affects the Cal-Alert
screen. It is not necessary to restart the instrument to activate the security setting.

4.2.3. Default System Parameters

Setup
PLC Remote
System
Single Step
Alarm
Contrast
Results
Lock
Mem Lock
Hipot Start
Date Format
Cal Alert
Alert Date
4.2.4. Memory, Step, and Default System Parameter Restore
Press the two bottom soft keys and power the instrument at the same time, then press
the soft key next to the word "Yes". All memories and steps will be loaded with the
ACW default parameters and the system parameters will be set to the factory defaults.
The following System parameters will not be affected by the system initialization:
Setup System
Cal Alert
Cal Date
Cal Due
Alert
Date m/d/y
Time

4.3. Test Setup

1. From the Perform Tests screen, press the "Menu" soft key. The Main Menu
screen will now be displayed.
____________________________________________________________________________
Restoring the instrument's System parameters will overwrite
all memories and steps with ACW default parameters!
OFF
OFF
5
5
Last
OFF
OFF
GW
mdy (month, day, year)
ON
Cal Date + 11 months
No change
No change
No change
No change
No change
No change
28