Technical Specifications
Item
Pace Annotation
Pace digital sampling rate
Table 124: ECG Data Processing Specifications
Item
ECG Interpretation
Computerized measurements
Heart rate meter
ECG data formats
Pre-acquisition
Digital rhythm
Storage
External storage
Downsampled ECG waveform
Analyzed ECG waveform
Additional report filters
Channels

Environmental Specifications

Table 125: Environmental Specifications
Item
Operating Conditions
Temperature
Relative Humidity (RH)
336
Description
Dedicated pace channel on display and printed reports
75,000 samples/second per channel
Description
Marquette 12SL ECG Analysis Program for Adults and Pediatrics
12-lead analysis
30 to 300 beats per minute (BPM) with an accuracy of ±10% or 5
BPM, whichever is greater. Heart rates outside this range will not
display.
GE Hi-Fidelity ECG, XML
Provides 10 seconds of instantaneous ECG acquisition
Up to 5 minutes of continuous rhythm storage (exportable as a
PDF)
300 records consisting of 10 second Resting ECG records and
Digital Rhythm records on the device internal memory
USB-compliant flash drive supporting the FAT32 file system
Bandwidth: 0.04 to 300 Hz
Sample rate: 2 ksps
Resolution: 1.22 µV
Bandwidth: 0.04, 0.56 ZPD to 300 Hz
Sample rate: 500 and 1000 sps
Resolution: 4.88 µV
20 Hz, 40 Hz, 100 Hz, 150 Hz or 300 Hz
Up to 12 channels, skew between channels: < 100 μS
Description
10ºC to 40ºC (50° F to 104° F)
20% to 95% (non-condensing)
  MAC
5 A4/MAC
5 A5/MAC
5 Lite
 5864335-001-1