GE MAC 5 A4 Operator's Manual page 203

Resting ecg analysis system
Table of Contents

Advertisement

Field
STAT Access
Label
Enable Login
Notification
Login
Notification
Message
User Password
Expiration
5864335-001-1
Action
Description
Enter the label
This field is enabled only when the Full Authentication
name to display in
with STAT setting is selected from the User
the login screen.
Authentication field.
Default value: STAT
Allowed values:
1 to 20 characters
A to Z
a to z
0 to 9
All special characters
Enable or disable
If this setting is enabled, the Login Notification
this setting.
Message field is enabled. You must configure a
notification message that displays at the time of login
and acknowledged by users who attempt to log in to
the device.
If this setting is disabled, the Login Notification
Message field is disabled.
Default value: Disabled
Enter the login
No default value
notification
Allowed values:
message.
Up to 15000 characters
A to Z
a to z
0 to 9
All special characters
Enable or disable
If this setting is enabled, set the duration for password
this setting.
expiration in the Password lifetime duration Minimum
and Maximum (days) fields. The password expires after
the configured duration, and the user is prompted to set
a new password.
If this setting is disabled, the password does not expire.
Default value: Disabled