Change the Order Attached to a Patient Test

If an incorrect order is attached to a patient test, use below procedure to detach the
order from the patient test and replace it with a different order.
Before you start this procedure, make sure that the patient test is not already
transmitted to its default destination. If the test has the Sent status of Yes (for
example, the test was already sent to its default destination), the order can no longer
be detached from the test. A message displays indicating that the patient test has
been transmitted and the order cannot be detached.
WARNING:
INACCURATE PATIENT DATA - Incorrect patient information can cause patient
data mismatch. Data assigned to the wrong patient causes erroneous patient
data that can affect diagnosis and treatment.
Be sure to check the patient information screen for each new patient. Make
sure that you attach the correct order for the correct patient.
1.
Detach the incorrect order from the patient test. See
Patient Test on page
2.
Attach the correct order to the patient test. See
Test is Completed on page

Detach an Order from a Patient Test

Before you start this procedure, make sure that the patient test with the incorrect
order has not been transmitted to its default destination.
NOTE:
1.
From the Files list, select the patient report with the incorrect order that you
want to detach.
2.
Expand the Patient Information screen and scroll down to the Order Number
field.
3.
Select Detach next to the Order Number field to clear the field.
104
opened from the Files list.
104.
If the patient test to which the order is attached has been transmitted to its
default destination, the order cannot be detached from the test. A message
displays indicating that the patient test has been transmitted and the order
cannot be detached.
  MAC
5 A4/MAC
Attach an Order when the Patient
102.
5 A5/MAC
5 Lite
Detach an Order from a
 5864335-001-1