GE MAC 5 A4 Operator's Manual page 131

Resting ecg analysis system
Table of Contents

Advertisement

Field
Hookup
Advisor Level
Print Preview
Mode
Acquisition
Mode
Delay Rhythm
Printing
5864335-001-1
Action
Description
Select a value
The Hookup Advisor indicator displays yellow or red
from the drop-
depending on the severity of the signal failure in a
down list to
leadwire.
configure when
If you select Yellow, the Electrode Placement Image
the Electrode
automatically expands when the indicator is yellow
Placement Image
or red. When the signal turns green, the Electrode
automatically
Placement Image automatically collapses.
expands, in
case of leadwire
If you select Red, the Electrode Placement Image
failures.
automatically expands when the indicator is red.
When the signal turns yellow or green, the Electrode
Placement Image automatically collapses.
If you select Never, the Electrode Placement
Image does not automatically expand or collapse,
regardless of the type of signal received. You
can manually expand and collapse the Electrode
Placement Image at any time.
Default value: Never
Select a value
If Always is selected, a preview of the ECG always
from the drop-
displays after ECG acquisition.
down list to
If Yellow is selected, a preview of the ECG displays
configure the
after ECG acquisition, if the Hookup Advisor status
print preview
for the 10-seconds ECG acquired is Yellow or Red.
mode.
If Red is selected, a preview of the ECG displays after
ECG acquisition, if the Hookup Advisor status for the
10-seconds ECG acquired is Red.
If Never is selected, a preview of the ECG is never
display after ECG acquisition.
Default value: Always
Select a value
If Pre-acquisition is selected, the system acquires
from the drop-
the latest/previous 10-seconds of data for analysis.
down list to
If Post-acquisition is selected, the system displays
configure the
the acquisition progress until 10-seconds of ECG
acquisition mode.
data is acquired.
Default value: Pre-acquisition
Enable or disable
If this setting is disabled, rhythm printing occurs in real-
this setting.
time.
If this setting is enabled, the system waits until 10-
seconds of rhythm data is acquired before starting
rhythm printing.
Default value: Disabled
  MAC
5 A4/MAC
5 A5/MAC
5 Lite
Configure Settings
 131