GE MAC 5 A4 Operator's Manual page 195

Resting ecg analysis system
Table of Contents

Advertisement

c)
Configure Order Settings
Make sure that order management is enabled. See
page
186.
1.
Select Settings > Workflow > Order Manager.
2.
Select Orders.
3.
Configure order settings as per the information in the table.
Table 58: Order Settings
Field
Allow
Unprivileged
User to Attach
Orders
Auto-Update
Show Pending
Orders
4.
Select Save.
Configure Columns for the Orders List
Make sure that order management is enabled. See
page
186.
The columns displayed in the Orders list on the Acquisition screen are configurable.
1.
Select Settings > Workflow > Order Manager.
2.
Select Columns.
5864335-001-1
Select Save.
Action
Enable or disable
this setting.
Enable or disable
this setting.
Enable or disable
this setting.
  MAC
5 A4/MAC
Configure Order Management on
Description
If this setting is enabled, any user who does not
have the privilege to view orders can search for
a matching order on the device or on the MUSE
system using the Patient ID or Visit Number, and
attach the order to the patient test. A patient query
mode must be configured and the user should be
assigned the Query Remote Patient Data privilege
for automatic patient query.
If this setting is disabled, users who do not have the
privilege to view orders cannot attach orders by
searching for them.
This does not apply to the default STAT user role.
Default value: Disabled
If this setting is enabled, the Orders list is automatically
updated from the configured order management server.
Default value: Enabled
If this setting is enabled, the system displays all
pending orders, regardless of the device used to
move the order to the pending state.
Default value: Enabled