Configure Critical Value Notifications

Before you start this procedure, make sure that:
• The CRIT option is enabled on the device. Contact a GE Healthcare Service Support
representative to enable this option.
• Your user role is assigned the privileges to access the Settings screen and edit
critical value settings. See
1.
Select Settings > ECG > Critical Values.
The Critical Values screen displays.
5864335-001-1
Action
Description
Select a value
This field is enabled only when Rhythm Mode is
from the drop-
configured as Digital Only or Both.
down list to
Default value: 300 sec
configure the
maximum digital
Allowed values: 10 Sec to 300 Sec in multiples of 10.
rhythm duration.
This field is enabled only when Rhythm Mode is
from the drop-
configured as Digital Only or Both.
down list to
Default value: 25.0 mm/s
configure the
speed at which
Allowed values:
the rhythm is
5.0 mm/s
recorded.
12.5 mm/s
25.0 mm/s
50.0 mm/s
Select a value
Default value: Letter
from the drop-
Allowed values:
down list to
configure the
A4 - available on MAC 5 A4 device
paper size for
Letter - available on MAC 5 A4 device
printing.
A5 - available on MAC 5 A5 device
Configure User Roles on page
  MAC
5 A4/MAC
5 A5/MAC
209.
5 Lite
Configure Settings