Configure 12SL Interpretations

1.
Select Settings > ECG > 12SL.
The 12SL screen displays.
5864335-001-1
Action
Description
Select a value
Gain indicates how many mm represent 1 mV of sample
from the drop-
data on the printout. Changing the gain changes the
down list, to
amplitude of the waveforms. A higher gain makes the
configure the
amplitude of the waveform appear higher; a lower gain
default gain of the
makes the amplitude of the waveform appear lower.
ECG waveform
The 10/5 mm/mV setting is used to display the limb
to display on
leads (I, II, III, aVr, aVI, and aVf) at 10mm/mV and chest
the Acquisition
leads (V1 - V6) at 5 mm/mV. This is sometimes done to
screen.
reduce or prevent waveform overlap in the chest leads,
while avoiding tiny waveforms in the limb leads.
The standard grid paper is divided into small squares of
1 mm x 1 mm and large squares of 5 mm x 5 mm. When
printing 10 mm/mV, 1 mV of data is represented in 10
mm (2 large squares) on the printout.
Default value: 10.0 mm/mV
Allowed values:
2.5 mm/mV
5.0 mm/mV
10.0 mm/mV
20.0 mm/mV
10.0/5.0 mm/mV
Select a value
A faster speed makes the waveform display further
from the drop-
apart; a slower speed makes the waveform display
down list, to
closer together.
configure the
The standard grid paper is divided into small squares of
default speed
1 mm x 1 mm and large squares of 5 mm x 5 mm. When
of the ECG
printing 25 mm/s, 1 second of data is represented in 25
waveform to
mm (5 large squares) on the printout.
display on the
Acquisition
Default value: 25.0 mm/s
screen.
Allowed values:
5.0 mm/s
12.5 mm/s
25.0 mm/s
50.0 mm/s
  MAC
5 A4/MAC
5 A5/MAC
5 Lite
Configure Settings
 135