Set Up The Dcp Server Configuration In The Muse System - GE MAC 5 A4 Operator's Manual

Resting ecg analysis system
Table of Contents

Advertisement

If it is not selected, select it now.
When selecting DCP, both DCP inbound and DCP outbound are also selected.
7.
Continue to click Next on each window until you reach the Choose MUSE
Options window.
8.
Verify that the DCP communication option is selected.
If it is not selected, select it now.
9.
Click Next.
The MUSE Serial Number window opens.
• If you added the DCP communication option in the previous step, you need to
• If you do not know the password, contact GE Healthcare Technical Support.
10. Click Next through the remaining screens until your changes are applied and the
Maintenance Complete window opens.
11. Click Finish.
NOTE:
12. Verify the installation.
See

Set Up the DCP Server Configuration in the MUSE System

By default, the DCP Inbound service has a Device Friendly Name of MUSE and listens
on port 9240 of all network interfaces on the MUSE application server.
Perform the steps below to modify these defaults:
1.
Log on to the MUSE system as a user with privileges to modify settings in MUSE
Setup.
2.
Go to System > Setup.
3.
In the Navigation pane, select System.
4.
Right-click on the MUSE entry and choose Properties.
The System Properties window opens.
5.
Select DCP Configuration.
6.
Modify the fields using the information in the table below.
5864335-001-1
enter the Options Configuration Password.
When communicating with a MUSE system via DCP communication, a
modem cannot be defined for the MAC system.
System Checkout on page
  MAC
Configure the MUSE System for Network Communication
332.
5 A4/MAC
5 A5/MAC
5 Lite
 329