GE MAC 5 A4 Operator's Manual page 233

Resting ecg analysis system
Table of Contents

Advertisement

Field
Probe Delay
Frequency
Band
Threshold (dB)
If the configured authentication protocol is:
• WEP, go to step 3.
• WPA or WPA2, go to step 4.
• Open, go to step 7.
3.
Configure WEP authentication as per the information in the table, and then go to
step 7.
NOTE:
5864335-001-1
Action
Enter the number
of seconds for
probe delay.
Select a value
from the drop-
down list to
configure the
frequency band
of wireless
operation.
Select a value
from the drop-
down list to
configure the
signal threshold in
dB.
You can secure some communication channels with encryption and
authentication. GE Healthcare recommends that you use the encrypted
channels, not the unencrypted channels. For example, WPA2 for wireless
authentication protocol instead of WEP.
  MAC
5 A4/MAC
Description
When the timer for this delay starts, the device checks
if wireless is enabled and the wireless network is
connected. If it is disconnected, the device will try to
reconnect to the wireless network.
Default value: 5
Allowed values: 5 to 120
Default value: Dual
Allowed values:
Dual
2.4 GHz
5 GHz
To enable the device to roam more frequently, decrease
the signal threshold.
To prevent the device from roaming frequently, increase
the signal threshold.
Default value: 80
Allowed values: 50, 55, 60, 65, 70, 75, 80, 85 or 90
5 A5/MAC
5 Lite
Configure Settings
 233

Advertisement

Table of Contents
loading

This manual is also suitable for:

Mac 5 a5Mac 5 lite

Table of Contents