Configure Filters, Gain, And Speed - GE MAC 5 A4 Operator's Manual

Resting ecg analysis system
Table of Contents

Advertisement

Configure Settings
Field
Simulated
Patient Data
Auto-ECG
Print Preview
Auto Full
Screen
3.
Click Save.

Configure Filters, Gain, and Speed

1.
Select Settings > ECG > Filters, Gain, Speed.
The Filters, Gain, Speed screen displays.
132
Action
Description
Enable or disable
If this setting is enabled, you can use simulation patient
this setting.
data for demonstrations or troubleshooting.
The system generates and displays simulated ECG
waveforms on the Acquisition screen. The label at the
top of the screen indicates that the ECG waveform is
based on simulated data from the internal simulator,
and not actual patient data.
If this setting is disabled, the system displays
waveforms recorded from a patient attached to the
device.
Default value: Disabled
Enable or disable
If this setting is enabled, as soon as the ECG signal is
this setting.
good, the device automatically starts recording 10-
seconds of ECG data for only one ECG per patient
connection.
Default value: Enabled
Enable or disable
If this setting is enabled, the system automatically
this setting.
displays the ECG preview window in full screen mode.
If this setting is disabled, the system displays the ECG
preview window in normal mode.
This setting is disabled, if the Print Preview Mode value
is configured to Never.
Default value: Enabled
  MAC
5 A4/MAC
5 A5/MAC
5 Lite
 5864335-001-1