Clean And Disinfect The Device And Trolley; Pre-Clean Inspection And Functional Test; Clean The Device And Trolley - GE MAC 5 A4 Operator's Manual

Resting ecg analysis system
Table of Contents

Advertisement

Clean and Disinfect the Device and Trolley

If you purchase a trolley, the device and trolley are designed to require regular
inspection and cleaning to function properly. The cleaning instructions for the device
includes the touchscreen.
WARNING:
ELECTRICAL HAZARD - Improper handling during inspection or cleaning could
result in electrical shock.
To avoid potential shock, observe the guidelines at all times:
Before inspecting or cleaning the device, turn it off, unplug it from AC power,
and remove the battery.
Do not immerse any part of the equipment in water.

Pre-Clean Inspection and Functional Test

Perform a visual inspection to verify that the device meets the minimum conditions:
• The case and display screen are free of cracks and other damage.
• All plugs, cords, cables, and connectors are free of kinks, frays, and other damage.
• All cords and connectors are securely seated.
• All keys and controls operate properly.
• The trolley exterior is free of cracks and other damage.
• The accessory track is functioning properly.
• All cords and connectors are securely seated.
• The actuation lever is functioning properly.
• The castor wheels are functioning properly.
If you notice any items that need repair, contact an authorized service representative
to make the repairs. Discontinue using the device until the appropriate repairs can be
made.

Clean the Device and Trolley

1.
Dispense Super Sani-Cloth
2.
If soil is present, thoroughly wipe the surfaces of the device with a fresh Super
Sani-Cloth
have been visibly removed.
Treated surfaces must remain visibly wet for a minimum of two minutes. Use
additional fresh disinfectant wipes, if needed, to make sure continuous two
minutes contact time. Pay attention to the recessed areas and ridges; use a
cotton swab to press onto the wipe for scrubbing these areas.
5864335-001-1
®
Wipe(s) from the canister.
®
Wipe for a minimum of two minutes and until soil and organic matter
  MAC
5 A4/MAC
5 A5/MAC
5 Lite
Cleaning and Disinfection