Adjustments / Programs
5-

Generator, kV offset adjustment

5Adjustments / Programs
NOTE
Connect the service PC to the serial interface of the SIREMOBIL Iso-C.
Start the service program.
Select the Diagnostic menu.
Select the Monitoring menu.
Select "KV offset adjustment" in the "Group 1" Combo box.
WARNING
Click "Update Start".
X
After the "Start fluoro" message in the ACTIONS line, press the radiation release but-
ton on the SIREMOBIL Iso-C hand switch.
NOTE
Continue to hold the radiation release button as long as the message "Continue fluoro"
is displayed in the ACTIONS line.
Release the radiation release button as soon as the message "Stop monitoring"
appears in the ACTIONS line.
Click "Update Stop".
Perform the generator adjustment (filament circuit learning program).
© Siemens, 2008
For internal use only
The kV offset adjustment has to be followed by the generator
adjustment (filament circuit learning program). Always per-
form the kV offset adjustment prior to beginning the genera-
tor adjustment.
X-ray radiation!
See
(Safety information / p.
Radiation protection is required. Wear a lead apron.
After the "KV error break" message in the STATUS line, click
"Update Stop". Wait approximately 10 seconds.
"Update Start" again. After the "Start fluoro" message in the
ACTIONS line, press the radiation release button on the
SIREMOBIL Iso-C hand switch. Refer to the text that follows
for additional procedures.
SPR2-230.840.01.01.02
08.08
11).