Dose-area product measurement device

Remove the SIREPHOS cover.
Replace the dose-area product measurement device.
Only for dose-area product measurement device from PTW:
- Perform the calibration of the dose-area product measurement device. See
(Dose-area product measurement device "PTW" / p.
Only for dose-area product measurement device from WELLHÖFER
- Set the slide switch on the dose-area product measurement device to position
"1 cGycm per pulse / test value 50" (fully to the right).
- See
(Dose-area product measurement device "WELLHÖFER" / p.
Test the accuracy of the dose-area product measurement device. See
accuracy of the dose-area product measurement device / p.
Reinstall the SIREPHOS cover.
Seal the SIREPHOS cover using sealing compound. See
SIREMOBIL Iso-C
Replacing boards and components
SPR2-230.840.01.01.02
08.08
65).
70).
(Covers / p.