Replacing boards and components

MULTISPOT 2000

Before installing the new MULTISPOT 2000, set the correct refresh rate. See Table 1.
50 Hz refresh rate
Jumper
X4
X5
X6
X7
X8
X9
X10
X11
Disconnect the SIREMOBIL line voltage plug from the receptacle.
Replace the MULTISPOT 2000.
The screws securing the MULTISPOT 2000 are accessible on the underside of the
camera.
Perform a camera function test after the replacement.
Generate and program the exposure parameters for the Multispot 2000 according to
the IQ Quick test (see the log book). Record the new exposure parameters in the IQ
Quick test instructions.
© Siemens, 2008
For internal use only
Position
Jumper
b
open
b
b
b
b
b
b
SPR2-230.840.01.01.02
08.08
60 Hz refresh rate
Position
X4
a
X5
open
X6
a
X7
a
X8
a
X9
a
X10
a
X11
a
Page 117 of 148
CS PS SP
117
6.17
SIREMOBIL Iso-C