Generator Adjustment

2Adjustments
1. Register an emergency patient to allow radiation release.
2. Place 2.1 mm Cu on the radiation exit port of the single tank.
3. Click on <Go>.
X
4. Start fluoro
5. Stop fluoro
X
6. Start fluoro
7. Stop fluoro
8. Click on <Save>.
9. Click on ">" for the next page (multiple exposure) and perform a generator adjustment
check.
ARCADIS Varic / Orbic
Action bar: "start fluoroscopy"
Action bar: "continue fluoro"
Status bar: "kV- offset-learning"
Status bar: "kV- learning done"
Action bar: "stop fluoroscopy"
Status bar: "generator learning"
Action bar: "start fluoroscopy"
Action bar: "continue fluoro"
Status bar: "warmup" (~ 8 minutes)
Status bar: "filament correction"
Status bar: "learning pushfactor"
Status bar: "gen. learning done"
Action bar: "stop fluoroscopy"
A window appears: "Main system values were successfully saved" --> click on
<OK>.