Siemens SIREMOBIL Iso-C Troubleshooting Manual page 69

Table of Contents

Advertisement

Adjustments / Programs
In the "Available Functions" window select the "Start Diamentor Test" line and click
"Execute".
Close the popup window by clicking the OK button.
Close the prompt "Do you want to reset the unit?" by clicking the "No" button.
Start the test by pressing the ->0<- key (reset key for the fluoroscopic time).
The tested value is displayed on the SIREMOBIL monitor.
The displayed value must agree with the programmed test value "50". Tolerance: 5 %.
If the displayed value is outside tolerance, replace the dose-area product measure-
ment device.
After the check on the service PC, select "Stop Diamentor Test" and click "Execute".
Close the popup window by clicking the OK button.
Close the prompt "Do you want to reset the unit?" by clicking the "No" button.
The test is finished.
Check the accuracy of the dose-area product measurement device. See
accuracy of the dose-area product measurement device / p.