Generator Adjustment (Filament Circuit Learning Program) - Siemens SIREMOBIL Compact Service Instructions Manual

Table of Contents

Advertisement

5 - 2

Generator adjustment (filament circuit learning program)

NOTE
NOTE
NOTE
Connect the service PC to the serial interface on board D1.
Start the service program.
Select the Diagnostic menu.
Select the Monitoring menu.
In the "Group 1" Combo box, select "Generator adjustment".
Click on "Update Start".
WARNING
After the "Start fluoro" message in the ACTIONS line, press the radiation release button
on the SIREMOBIL Compact hand switch.
Continue to hold the radiation release button as long as the message "Continue fluoro" is
displayed in the ACTIONS line.
"Warmup" appears in the STATUS line, the Sirephos is warming up. After approx. 8
minutes, the message "learning pushfactor" will appear in the STATUS line.
A maximum of 15 exposures will automatically be released.
The message "Learning done" appears in the STATUS line after the generator
adjustment has been successfully completed.
The "Stop monitoring" message appears in the ACTIONS line.
Stop radiation.
Click on "Update Stop".
Switch the SIREMOBIL Compact OFF and back ON again.
SIREMOBIL Compact
System Manual
The kV offset adjustment must always be performed prior to start-
ing the generator adjustment.
If an error occurs during the generator adjustment, switch OFF
the SIREMOBIL Compact and perform the kV offset and generator
adjustments again.
As soon as the generator adjustment is started, press the radia-
tion release button on the hand switch until the "Stop monitoring"
message appears in the ACTIONS line. If you deactivate radiation
too soon, the generator adjustment must be completely repeated.
Exception: You can briefly stop pressing the radiation release
button (less than 30 seconds) during the warmup phase (warmup
displays in the STATUS line on the service PC) of the SIREPHOS.
The generator adjustment continues when the radiation release
button is pressed again.
X-ray radiation!
See Chapter 1, Safety Information.
Protect yourself against radiation exposure. Wear a lead apron.
Register 5
RXR2-130.061.02
Rev. 03
Adjustments / Programs
Page 2 of 22
10.03
CS PS 24
5
Siemens AG
Medical Engineering

Advertisement

Table of Contents
loading

Table of Contents