12
WARNING
WARNING
WARNING
X
SIREMOBIL Iso-C
Electrical voltage!
Noncompliance can lead to death, physical injury or property
damage.
After completing service work and reattaching all cov-
ers, a protective conductor test must be performed
according to ARTD-002.731.17.. . The protective con-
ductor resistance must not exceed 0.2 ohms.
Danger of infection with pathogens!
Noncompliance can lead to death or physical injury.
This product has been released for operation in operat-
ing rooms, and may therefore be contaminated with
infectious blood or other bodily fluids.
Any skin contact with blood or other bodily fluids must
be avoided! For this reason, you must strictly adhere
to the preventive measures specified in
ARTD-002.731.37... !
X-ray radiation!
Noncompliance can lead to illness, irreversible damage to
body cells and the genotype, severe injury and even death.
In the case of system work requiring the release of radiation,
the radiation protection guidelines and the rules for radiation
protection according to ARTD-002.731.02 must be complied
with.
Note:
Use available radiation protection devices.
Wear radiation protection clothing (lead apron).
Maintain the greatest possible distance from the radia-
tion source.
Release radiation only if necessary.
Set the radiation activity as low as possible (low kV and
mA values, short radiation time).
Release radiation for as short a time as possible.
Checks requiring the release of radiation are identified
by the radiation warning symbol.
SPR2-230.840.01.01.02
08.08
Page 12 of 148