Checking The Generator Adjustment - Siemens SIREMOBIL Iso-C Service Instructions Manual

Hide thumbs Also See for SIREMOBIL Iso-C:
Table of Contents

Advertisement

Adjustments / Programs
Switch the SIREMOBIL Iso-C OFF and back ON again.

Checking the generator adjustment

With the system on, connect the service PC to the SIREMOBIL Iso-C.
Select "Test..." in the Diagnostic menu.
Select cassette simulation in "Function Groups" .
Select cassette switch ON in "Available Functions". The SIREMOBIL Iso-C is now in di-
rect exposure mode.
Select 81 kV with the kV+ / kV- keys.
Select 8 mAs with the mAs+ / mAs- keys.
Connect oscilloscope channel 1 to D1.X46, ground at 0V.
Set the sensitivity to 1V / Div.
Connect oscilloscope channel 2 to D1.X48, ground at 0 V.
Set the sensitivity to 2.5V / Div.
Set the time to 25 ms / Div.
Set the oscilloscope on storage mode.
Release an exposure with the hand switch and store the oscillogram with the exposure.
Evaluate the stored oscillograms.
The tube current must reach the tolerance range of 13 mA + 1.5 mA in the time of 3 ms
and may not leave this range for the next 6 ms.
(The tube current may not run above 14.5 mA within the first 6 ms).
Refer also to oscillograms in Fig. 1 / Fig. 1a and Fig. 1b.
Select "Cassette switch off" on the service PC in "Available Functions". Direct exposure
mode is cancelled.
If the unit exceeds the tolerance range, the generator adjustment must be repeated.
Siemens AG
Medical Solutions
SPR2-230.061.01
Rev. 13
12.04
Page 3 of 30
SIREMOBIL Iso-C
CS PS 24
5 - 3
5
System Manual

Advertisement

Table of Contents
loading