Replacing boards and components
Retighten the Allen screw of the large pulley.
Perform the mechanical settings as described in the Startup Instructions, Chapter 6.
To avoid damage to the new Orbital Pot when replacing, just do this:
Put new pot in but do not tighten the grub screw. Go into main
system<>Poti adjust and move the pot to give a reading of 200 when C
Bow it set at 95', then tighten grub screw..