Adjustments / Programs
Close the status window by clicking the OK button.
Respond to the query "Do you want to reset the unit" by clicking "No".
X
Switch on fluoro.
Open the slot diaphragm to where the blades begin to disappear at the edge of the
image.
Switch off fluoro.
In the window "Available Functions:" select "FULL FORMAT POSITION" and click
"Execute".
Close the status window by clicking the OK button.
Respond to the query "Do you want to reset the unit" by clicking "No".
Select the zoom format.
X
Switch on fluoro.
Open the slot diaphragm to where the blades begin to disappear at the edge of the
image.
Switch off fluoro.
In the window "Available Functions:" select "ZOOM FORMAT POSITION" and click
"Execute".
Close the status window by clicking the OK button.
Respond to the query "Do you want to reset the unit" by clicking "No".
In the window "Available Functions:" select "Enable circle mask" and click "Execute"
Respond to the query "Do you want to reset the unit" by clicking "No".
In the window "Available Functions:" select "INIT OFF & STORE POSITION" and click
"Execute".
Close the status window by clicking the OK button.
Respond to the query "Do you want to reset the unit" by clicking "Yes". The adjustment
values are stored.

Concluding steps

Switch off the system and switch it back on again.
Return to the main menu by pressing ESC at the service PC.
Select the "Data, Backup..." menu.
Click "parameter" in the Backup window.
Then click "Backup".
Enter "slot diaphragm adjusted" under "Remarks for Parameter".
Then click "OK" . The new position values are stored to disk.
You can now exit the service program.
© Siemens, 2008
For internal use only
SPR2-230.840.01.01.02
08.08
Page 55 of 148