3D Reconstruction System Electronics - Siemens SIREMOBIL Iso-C Troubleshooting Manual

Table of Contents

Advertisement

Replacing boards and components

3D reconstruction system electronics

Fig. 39:
NOTE
The 3D reconstruction system electronics (PC boards D200 including MCB3 and D310)
is attached in front of PC board D1 in the basic unit
Switch the system off.
Unplug all connectors.
Replace the module.
Reconnect all connectors.
Fasten the shielded cables in the shield connectors again, ensuring good ground con-
tact of the shielding.
If cable ties have been removed, reattach the cables with cable ties.
Perform a download of the control software as described in the Startup Instructions,
Chapter 5.
Perform the mechanical settings as described in the Startup Instructions, Chapter 6.
Perform chapters 10 and 11 of the IQ quick test.
© Siemens, 2008
For internal use only
Only applies to SIREMOBIL Iso-C 3D.
SPR2-230.840.01.01.02
08.08
((Fig. 39 / p.
Page 125 of 148
CS PS SP
125
6.25
125)).