Dose-area product measurement device "WELLHÖFER"

Programming in the SIREMOBIL

Connect the service PC to the serial interface of the SIREMOBIL.
Start the service program.
Select the "Adjustment", "Parameters..." menu.
In the combo box "Parameter Groups:" select "Diamentor".
Click "Get from Unit".
The Parameter "1. Diamentor Existing" must be set to YES .
Select parameter "2. Constant (Qt * Kg)".
Enter the test value "50" in the "Value Actual:" field.
Click "Set Value".
Click "Put to Unit".
The system is automatically reset and restarted.
After the system has booted, close the popup window by clicking the OK button.
Click "Save to File".
NOTE

Functional test of the measurement device and cabling

NOTE
Prerequisites
The test value is already programmed. See
Connect the service PC to the serial interface of the SIREMOBIL.
Start the service program.
Select the "Adjustment", "Calibration..." menu.
At the prompt click "Yes".
In the combo box "Function Group:" select "Diamentor".
SIREMOBIL Iso-C
The service program must be exited with "Logoff" and "Quit"
. Only then are the changed parameters saved as a new file
on floppy disk.
The check of the programmed value activates the internal
test of the dose-area product measurement device electron-
ics unit.
SPR2-230.840.01.01.02
08.08
Adjustments / Programs
'(Programming in the SIREMOBIL / p.
Page 68 of 148
CS PS SP
5.12
0
0
68)'.