Vertical lift limit switches

Fig. 58:
Replace the limit switches. Reconnect the wires.
To adjust the limit switches, move the lifting column slowly and carefully by briefly tap-
ping the lifting column switch at the control console of the basic unit until it is positioned
just before the mechanical end stop.
Adjust the respective limit switch and tighten the screws.
Make sure the limit switch is functioning correctly.
To do this, move back out of the end position.
Then slowly and carefully move to the end position by briefly tapping the lifting column
switch on the control console of the basic unit.
The lifting column motor must switch off prior to reaching the end position and the lifting
column must come to a full stop.
Move the lifting column approximately 10 cm away from the end position.
Now move the lifting column at full speed in the direction of the end position by pressing
and holding the lifting column switch.
The lifting column motor must switch off prior to reaching the mechanical end stop, and
the lifting column must come to a full stop.

Switch for blocking vertical lift

Remove the horizontal lift cover to replace both switches. See
cover / p.
Adjustment: Refer to the MODULARIS service instructions.

Emergency off switch

Loosen the horizontal carriage cover.
Unsolder the emergency off switch.
SIREMOBIL Iso-C
20).
SPR2-230.840.01.01.02
08.08
Brakes / Movements
(Horizontal carriage
Page 142 of 148
CS PS SP
7.5
0
0
© Siemens, 2008