Siemens SIREMOBIL Iso-C Troubleshooting Manual page 112

Table of Contents

Advertisement

112
Number of Monitors
Hardcopy
Edge 1 coef set
Edge 2 coef set
Motion detection setup
Date / Time format
Keyboard
Language
Display of realtime clock
Display of rotated images
Monitor zoom
Monitor split
Monitor text
Maximum image number
Disable dose rate selection
in user setup
Unit serial number
Default LUT
Default Edge
Software backward com-
patible
Product name
Default program change-
able
Create external disk
SIREMOBIL Iso-C
Replacing boards and components
Configure 1 for SIREMOBIL with 1 monitor. Configure 2 for
SIREMOBIL with 2 monitors.
Check whether 5000 is programmed in "4 Max Hardcopy
time". If not, program this value.
Check whether 3 is programmed; if not, program this value.
Check whether 8 is programmed, If not, program this value.
Check whether the following are programmed "MD1:
Threshold" 100 and for "MD2: Threshold" 120. If not, pro-
gram these values.
Set the country-specific format.
Set to the keyboard layout being used.
Set for language of the country.
Configure Yes. Select No only upon customer request.
Configure Yes.
Configure No.
Configure Yes.
Configure Yes.
Configure 700.
Set to "No" if equipped with an dose-area product measure-
ment device. The menu "Storage of Dose on disk" and the
menu "Dose on Hardcopy output" are visible in User Setup.
Otherwise set to "Yes".
Program the serial number of the SIREMOBIL. Do not enter
the leading zero.
Configure 2. Only program a different LUT curve upon cus-
tomer request.
Configure 1. Program a different edge enhancement value
upon customer request only.
Configure No.
Program "SIREMOBIL Iso-C".
Default: Yes. Depending on country-specific regulations
configure Yes or No.
If the option "3D Reconstruction" or the option "DICOM
Bridge" is installed or a MOD drive is present, configure
Yes. Otherwise configure No.
SPR2-230.840.01.01.02
08.08
Page 112 of 148
CS PS SP
© Siemens, 2008
For internal use only