Adjustments / Programs
Select "5. CLOSED POSITION" on the service PC and click "Execute".
Close the status window by clicking the OK button.
For the query "Do you want to reset the Unit", click "No".
Select "8. Init Off & Store Positions" on the service PC and click "Execute".
Close the status window by clicking the OK button.
For the query "Do you want to reset the Unit", click "Yes".
Switch the SIREMOBIL Iso-C off and on.
X
Check the X-Iris again in survey format and zoom format. The collimator blades must
still be visible on the edge of the image. If necessary, repeat the adjustment.
Return to the main menu on the service PC with the ESC key.
Select the Data, Backup... menu.
Select Parameter in the backup window.
Click "Backup".
At "Remarks for Parameter" enter "X-Iris set".
Click "OK" . The new position values for the collimator will be stored on the disk.
Return to the main menu with the ESC key.
© Siemens, 2008
For internal use only
SPR2-230.840.01.01.02
08.08
Page 53 of 148
CS PS SP
53