Collimator - Siemens SIREMOBIL Iso-C Troubleshooting Manual

Table of Contents

Advertisement

92

Collimator

Replace the collimator.
Ensure that the collimator is functioning properly and is set correctly. If necessary,
adjust it.
X
Ensure that the collimator blades of the iris diaphragm are visible during fluoroscopy in
survey format and zoom format on at least 2 sides of the monitor image.
Check the collimation for direct exposure if a cassette holder is being used.
Check the display of the blades on the monitor and adjust if necessary.
Remove the old set of collimator labels near the SIREPHOS and attach the new labels
(included in delivery) in the same position. Check that the original collimator labels are
the same as the new ones.
If the "MODULARIS" system is present, perform the isocenter check.
SIREMOBIL Iso-C
Replacing boards and components
SPR2-230.840.01.01.02
08.08
Page 92 of 148