Vertical Adjustment - Siemens MAMMOMAT Novation DR Installation Manual

Hide thumbs Also See for MAMMOMAT Novation DR:
Table of Contents

Advertisement

5 - 6
Fig. 7 Display at the lower part of the stand
1.
2.
NOTE

Vertical adjustment

CAUTION
1.
DR
MAMMOMAT Novation
System
Set the projection angle to 60 degrees with the presetting knob (1/Fig. 6).
The preset angle is displayed for approximately 3 s.
Press one of the buttons (2/Fig. 6).
The rotary motion of the swivel arm system shall stop automatically when the
preset projection angle (both + and -) is reached and at 0 degrees.
All eight rotary motion buttons (four on either side 2/Fig. 6) must
be checked. After the button has been released, the swivel arm
system shall remain in the position set.
Remove the protective strips before performing the vertical
adjustment of the swivel arm.
If not removed, the protective strips could be damaged.
Make sure the protective strips are removed before adjustments,
see "Metal Curtain" on Page 2 - 2.
Press the buttons for upward and downward movement (3/Fig. 6) of the swivel
arm system.
SPB7-250.815.02
Rev. 02
09.04
Mechanical Checks
Page 6 of 12
CS SD 24
5
Siemens AG
Medical Solutions