Overview
12
1.4

Automatic sensor recognition

The automatic sensor recognition for IDS sensors allows
to operate an IDS sensor at different meters without recalibration
to operate different IDS sensors at one meter without recalibration
to assign measurement data to an IDS sensor
– Measurement datasets are always stored and output with the
sensor name and sensor series number.
to assign calibration data to an IDS sensor
– Calibration data and calibration history are always stored and
output with the sensor name and sensor series number.
to activate the correct cell constant for conductivity sensors auto-
matically
to hide menus automatically that do not concern this sensor
To be able to use the automatic sensor recognition, a meter that sup-
ports the automatic sensor recognition (e.g. Multi 3410) and a digital
IDS sensor are required.
In digital IDS sensors, sensor data are stored that clearly identify the
sensor.
The sensor data are automatically taken over by the meter.
1.5

IDS sensors

IDS sensors
support the automatic sensor recognition
show only the settings relevant to the specific sensor in the setting
menu
process signals in the sensor digitally so that precise and interfer-
ence-free measurements are enabled even with long cables
facilitate to assign a sensor to a measured parameter with differently
colored couplings
have quick-lock couplings with which to fix the sensors to the meter.
Note
Information on available IDS sensors is given on the Internet and in
the WTW catalog, "Laboratory and field instrumentation".