wtw Multi 3410 Operating Manual page 63

Digital ph / dissolved oxygen / conductivity meter for a digital ids sensor
Hide thumbs Also See for Multi 3410:
Table of Contents

Advertisement

Multi 3410
Criteria for a stable
measured value
ba75863e02
12/2009
2
Using <MENU/ENTER>, activate the Stability control function
manually.
The [AR] status indicator appears while the measured value is
assessed as not stable. A progress bar is displayed and the
display of the measured parameter flashes.
The [HOLD][AR] status indicator appears as soon as a stable
measured value is recognized. The progress bar disappears
and the display of the measured parameter stops flashing.
The current measurement data is output to the interface. Mea-
surement data meeting the stability control criterion is marked
by AR.
Note
You can prematurely terminate the Stability control function manually
with <MENU/ENTER> at any time. If the Stability control function is pre-
maturely terminated, the current measurement data are not output to
the interface.
3
Using <MENU/ENTER>, start a further measurement with Sta-
bility control.
or
Release the frozen measured value again with <AR>.
The display switches to the measured value display.
The [AR][HOLD] status display disappears.
The Stability control function checks whether the measured values are
stable within the monitored time interval.
Measured
Time interval
parameter
ӕ
Conductivity
10 seconds
Temperature
10 seconds
The minimum duration until a measured value is assessed as stable is
the monitored time interval. The actual duration is mostly longer.
Stability in the time interval
ӕ
Δ
: better than
0.5% of measured value
Δ T (° C): Better than 0.02
Operation
63