wtw Multi 3410 Operating Manual page 44

Digital ph / dissolved oxygen / conductivity meter for a digital ids sensor
Hide thumbs Also See for Multi 3410:
Table of Contents

Advertisement

Operation
Sample record
Calibration evaluation
44
31.07.2009 15:55
Multi 3410
Ser. no. B211371282
CALIBRATIONpH
31.07.2009
16:13:33
SenTix 940
Ser. no. B211371282
AutoCal TEC
Buffer 1
Buffer 2
Buffer 3
Voltage 1
Voltage 2
Voltage 3
Slope
Asymmetry
Sensor
etc...
After calibrating, the meter automatically evaluates the calibration. The
zero point and slope are evaluated separately. The worse evaluation of
both is taken into account. The evaluation appears on the display and
in the calibration record.
Display
Clean the IDS sensor according to the
sensor operating manual
Error
Eliminate the error according to
chapter 6 W
HAT TO DO IF
Note
For pH IDS sensors you can optionally enable a more finely graded cal-
ibration evaluation (QSC) (see section 4.6.10).
4.01
7.00
10.01
184.0 mV
24.0 °C
3.0 mV
24.0 °C
-177.0 mV
24.0 °C
-60.2 mV/pH
4.0 mV
+++
Calibration
Zero point
record
[mV]
+++
-15 ... +15
++
-20 ... +20
+
-25 ... +25
-
-30 ... +30
< -30 or
Error
> 30
...
Multi 3410
Slope
[mV/pH]
-60.5 ... -58
-58 ... -57
-61 ... -60.5
or
-57 ... -56
-62 ... -61
or
-56 ... -50
... -62 or
... -50
ba75863e02
12/2009