1.3.2 IDS adapter for analog sensors

With the aid of an IDS adapter, you can also operate analog sensors on the
Multi 9310. The combination of the IDS adapter and analog sensor behaves
like an IDS sensor.
The measuring electronics with the stored adapter data are in the adapter
head. The adapter data correspond to the sensor data.
Information on available IDS adapters is given on the Internet and
in the WTW catalog, "Laboratory and field instrumentation".
Detailed information on the IDS adapter is given in the operating
manual of the adapter.

1.3.3 Automatic sensor recognition

The automatic sensor recognition for IDS sensors allows
to operate an IDS sensor with different meters without recalibrating
to assign measurement data to an IDS sensor
– Measurement datasets are always stored and output with the sensor
name and sensor series number.
to assign calibration data to an IDS sensor
– Calibration data and calibration history are always stored and output with
the sensor name and sensor series number.
to hide menus automatically that do not concern this sensor
To be able to use the automatic sensor recognition, a meter that supports the
automatic sensor recognition (e.g. Multi 9310) and a digital IDS sensor are
required.
In digital IDS sensors, sensor data are stored that clearly identify the sensor.
The sensor data are automatically taken over by the meter.