wtw Multi 3410 Operating Manual page 48

Digital ph / dissolved oxygen / conductivity meter for a digital ids sensor
Hide thumbs Also See for Multi 3410:
Table of Contents

Advertisement

Operation
Continuing with three-
point to five-point
calibration
(Buffer TEC)
48
22.09.2009 08:00
10
Wait for the measurement with stability control to be completed
or terminate the stability control and take over the calibration
value with <MENU/ENTER>.
The calibration display for the next buffer appears.
22.09.2009 08:00
11
Finish the calibration procedure as a two-point calibration with
<M>.
The calibration record is displayed.
12
Thoroughly rinse the IDS pH sensor with distilled water.
13
Immerse the IDS pH sensor in buffer solution 3.
14
Start the measurement with <MENU/ENTER>.
The measured value is checked for stability (stability control).
The [AR] status indicator is displayed. The measured parame-
ter flashes.
Multi 3410
ba75863e02
12/2009