Carrying out a QSC
control calibration
ba75863e02
12/2009
4
As soon as the three-point calibration has been successfully
carried out you can decide whether to accept or discard the cal-
ibration as the QSC initial calibration.
The QSC initial calibration is completed. The sensor is calibrated. If you
want to calibrate with special buffers for your measurements, you can
subsequently carry out a normal calibration with your buffers. The ref-
erence values determined with the QSC calibration are also used for
the evaluation of normal calibrations. In the measured value display,
the color scale of the QSC function is always displayed. A double arrow
on the color scale indicates the current sensor evaluation.
Info
22.09.2009 08:00
You can carry out QSC control calibrations at greater intervals than
normal calibrations.
A QSC control calibration can, e.g. be useful if the sensor evaluation
noticeably changed (after some normal calibrations).
1
Open the menu for measurement settings with <MENU/
ENTER>.
2
In the QSC menu, select Control calibration with < >< >.
The calibration display appears. AutoCal QSC-Kit is displayed
as the buffer.
Exclusively use the QSC-Kit for the QSC calibration. If you use
other buffers, you will have no valid QSC control calibration.
3
Follow the user guide.
The calibration is carried out like a normal three-point calibra-
tion. As soon as the three-point calibration has been success-
fully carried out you can decide whether to accept or discard the
calibration as the QSC control calibration.