wtw Multi 3410 Operating Manual page 47

Digital ph / dissolved oxygen / conductivity meter for a digital ids sensor
Hide thumbs Also See for Multi 3410:
Table of Contents

Advertisement

Multi 3410
Continuing with two-
point calibration
(Buffer: TEC)
ba75863e02
12/2009
22.09.2009 08:00
5
Wait for the end of the measurement with stability control or
accept the calibration value with <MENU/ENTER>.
The calibration display for the next buffer appears.
22.09.2009 08:00
6
If necessary, finish the calibration procedure as a single-point
calibration with <M>.
The calibration record is displayed.
Note
For single-point calibration, the instrument uses the Nernst slope
(-59.2 mV/pH at 25 °C) and determines the zero point of the IDS pH
sensor.
7
Thoroughly rinse the IDS sensor with distilled water.
8
Immerse the IDS pH sensor in buffer solution 2.
9
Start the measurement with <MENU/ENTER>.
The measured value is checked for stability (stability control).
The [AR] status indicator is displayed. The measured parame-
ter flashes.
Operation
47