wtw Multi 3410 Operating Manual page 51

Digital ph / dissolved oxygen / conductivity meter for a digital ids sensor
Hide thumbs Also See for Multi 3410:
Table of Contents

Advertisement

Multi 3410
ba75863e02
12/2009
22.09.2009 08:00
6
Wait for the measurement with stability control to be com-
pleted.
Set the nominal buffer value for the measured temperature with
< >< >.
7
Take over the calibration value for the asymmetry with <MENU/
ENTER>.
The calibration display for the next buffer appears.
22.09.2009 08:00
8
If necessary, finish the calibration procedure as a single-point
calibration with <M>.
The calibration record is displayed.
Note
For single-point calibration, the instrument uses the Nernst slope
(-59.2 mV/pH at 25 °C) and determines the zero point of the IDS pH
sensor.
Operation
51