Operation
54
4.6.9 Continuous measurement control (CMC function)
The Continuous Measurement Control (CMC function) facilitates to
evaluate the current measured value instantly and definitely.
After each successful calibration the scale of the pH measuring range
is displayed in the measured value display. Here you can very clearly
see whether or not the current measured value is in the calibrated part
of the measuring range.
The following information is displayed:
Info
22.09.2009 08:00
1
Measuring range for which a valid calibration is available (back-
ground color). Measured values in this range are suitable for
documentation.
2
Measuring range for which no valid calibration is available (light
gray). Measured values in this range are not suitable for docu-
mentation. Calibrate the meter with buffers covering this mea-
suring range.
If the current measured value is not in the calibrated range, the
color of this range changes to dark gray.
If a measured value is outside the measuring range pH 0 - 14,
overflow arrows are displayed at the left or right edge of the mea-
suring range.
3
Currently measured pH value (needle)
4
Marking lines for all nominal buffer values used with the last valid
calibration
The limits of the calibrated range are determined by the buffers used
for calibration:
Lower limit:
Buffer with lowest pH value - 2 pH units
Upper limit:
Buffer with highest pH value + 2 pH units
Multi 3410
1
2
3
4
ba75863e02
12/2009