Displaying Calibration Records - wtw Multi 3410 Operating Manual

Digital ph / dissolved oxygen / conductivity meter for a digital ids sensor
Hide thumbs Also See for Multi 3410:
Table of Contents

Advertisement

Operation
Continuing with two-
point calibration
Displaying the
calibration record
52
9
Thoroughly rinse the IDS pH sensor with distilled water.
10
Immerse the IDS pH sensor in buffer solution 2.
11
Start the measurement with <MENU/ENTER>.
The measured value is checked for stability (stability control).
The measured parameter flashes.
12
Wait for the measurement with stability control to be com-
pleted.
Set the nominal buffer value for the measured temperature with
< >< >.
22.09.2009 08:00
13
Take over the calibration value for the slope with <MENU/
ENTER>.
The calibration is completed as a two-point calibration.
The calibration record is displayed.

4.6.8 Displaying calibration records

The calibration data can be displayed and then output to the interface.
The calibration record of the last calibration is to be found under the
menu item, Calibration / Calibration record. To open it in the measured
value display, press the <CAL
The calibration records of the last 10 calibrations are to be found in the
menu, Calibration / Calibration data storage. To open the Calibration
menu, press the <MENU/ENTER> key in the measured value display.
_
> key.
ba75863e02
Multi 3410
12/2009

Advertisement

Table of Contents
loading

Table of Contents