wtw Multi 3410 Operating Manual page 67

Digital ph / dissolved oxygen / conductivity meter for a digital ids sensor
Hide thumbs Also See for Multi 3410:
Table of Contents

Advertisement

Multi 3410
Setting menu of
LR325/01
ba75863e02
12/2009
Menu item
Measurement /
Measuring cell /
Cell constant
Measurement /
Temp. comp. (TC) /
Method
Measurement /
Temp. comp. (TC) /
Linear coeff.
Measurement /
Temp. comp. (TC) /
Reference temp.
Measurement /TDS
factor
Reset
Possible set-
Description
ting
0.090 ... ...
Display and setting option of
0.100 ...
the cell constant of any mea-
-1
0.110 cm
suring cells (man).
nLF
Procedure for temperature
lin
compensation (see section
Off
4.8.3).
This setting is only available
for the measured parame-
ters,
0.000 ...
Coefficient of the linear tem-
2.000 ...
perature compensation.
3.000 %/K
This setting is only available
when the linear temperature
compensation is set.
20 °C
Reference temperature
25 °C
This setting is only available
for the measured parame-
ters,
0.40 ... 1.00
Factor for TDS value
-
Resets all sensor settings to
the delivery condition (see
section 4.11.1).
Operation
ӕ
ρ
and
.
ӕ
ρ
and
.
67