Omron E5GC User Manual page 356

Digital temperature controllers
Hide thumbs Also See for E5GC:
Table of Contents

Advertisement

7 User Calibration
adj
30
• Input type 0:
p390
e20c
• Input types 1, 2, 3, 4:
p280
e26b
p 10
4543
str
no
7 - 8
3. Turn the power ON.
4. Move to the Calibration Level.
This starts the 30-minute aging timer. This timer provides an approximate timer for
aging. After 30 minutes have elapsed, the No. 2 display changes to 0. You can
advance to the next step in this procedure even if 0 is not displayed.
5. Execute calibration for the main input.
Press the M Key to display the count value for each input type.
The No. 2 display at this time shows the currently entered count value in hexadecimal.
Set the 6-dial as follows:
Input type 0:
Input type 1, 2, 3 or 4:
Allow the count value on the No. 2 display to fully stabilize, then press the D Key to
temporarily register the calibration settings.
If this count value is outside of the specified range, the No. 2 display will flash and the
count value will not be temporarily registered.
6. When the M Key is pressed, the status changes as shown to the left.
Set the 6-dial to 10 Ω.
Allow the count value on the No. 2 display to fully stabilize, then press the D Key to
temporarily register the calibration settings.
If this count value is outside of the specified range, the No. 2 display will flash and the
count value will not be temporarily registered.
7. When the M Key is pressed, the status changes as shown to the left.
The data to be temporarily registered is not displayed if it is not complete.
Press the U Key. The No. 2 display changes to yes. Release the key and wait two
seconds or press the M Key. This stores the temporarily registered calibration data to
non-volatile memory.
To cancel the saving of temporarily registered calibration data to non-volatile memory,
press the M Key (while no is displayed in the No. 2 display) without pressing the U
Key.
8. The calibration mode is quit by turning the power OFF.
For Digital Controllers that have a transfer output, you can continue by calibrating the
transfer output. For detailed setting methods, refer to 7-6 Calibrating the Transfer
Output.
390 Ω
280 Ω
E5@C Digital Temperature Controllers User's Manual (H174)