[Options]
→ Identification: Information about manufacturer
and type of automatic furnace
[Options]
→ Statistics: Display of error statistics.
>
Automatic furnaces are equipped with a circular buffer
memory: Error messages are overwritten when the error log
is full. The fault that has occurred last is at position 1 in the
error list.
>
[Options]
→ Fault: Display of faults.
6.2.15.

Solid fuel measurement

The solid fuel measurement program is only available with
instruments that come with the CO sensor (not COlow). For
measurements in accordance with the 1st Federal Immission
Control Ordinance (BImSchV), a CO,H2-compensated sensor
(0393 0101) is required.
The function is only available if the activated fuel is solid
fuel and adapter 0600 9765 has been connected.
Calling up the function:
>
[
]
Measurement options
Performing the measurement:
1. Select parameters gas phase, meas. rate or stability time: [▲],
[▼]
→ [Change].
2. Enter values. [▲], [▼]and partly [◄],
3. [Finished].
4. Insert the flue gas probe into the exhaust pipe and position it in
the hot spot.
5. Start measurement:
-
The stabilisation phase (minimum 2 min) progresses. After this
the measuring phase starts automatically (minimum 5 min).
The stabilisation phase can be interrupted early.
> Press [Next].
-
The measuring phase starts automatically.
-
The measuring result is displayed after the measuring phase
has been completed.
www.
www.
6 Using the product
[OK]
Solid fuel
[►]
→ [OK].
[
]
information@itm.com
information@itm.com
.com
.com
→ [OK].
63